Improvement and FEA of New Type of Osteointegrated Implant For Higher Amputated Leg Attachment

文献类型: 会议论文

第一作者: J. M Luo

作者: J. M Luo 1 ; L. Zheng; X. H. Shi; Y. Wu; X.D. Zhang;

作者机构: 1.Engineering Research Centre in Biomaterials, Sichuan University, Chengdu, Sichuan, China

关键词: percutaneous cevice;stress distribution;stress concentration

会议名称: Advanced Biomaterials VII

主办单位:

页码: 829,832-

摘要: Stress concentration is one of the main mechanical problems leading to the failure of clinical application for osteointegrated implant of percutaneous osteointegrated prosthesis, which is especially marked for higher amputated leg prosthesis. Traditionally design was composed of only the distal part. To improve the biomechanical safety, a new design with the lag part similar to the lag screw was introduced. Based on CT scan data, relatively accurate model of femur for finite element analysis (FEA) were obtained. The FEA results with the new implant demonstrated that compared to traditional design, the declination of bone stress peak ranged from 15.68% to 28.67%, perpendicular deformation from 34.73% to 72.16%, and maximal stress of implant from 14.51% to 23.36% with the increasing of loads from 3750N to 2000N. So the new design of osteointegrated implant would be more secure mechanically, in the case of higher amputated leg attachment.